8   23-03-2009  105446  06 
 The ATC-125 contains the gases R-1270 and R-704 
under pressure. The calibrator must under no 
conditions be stored or operated at ambient 
temperatures above 60C (140F). Doing so may cause 
a hazard. 
About the front panel: 
 For B versions only, the sockets on the input module  
must NEVER be connected to voltages exceeding 5V 
for the TC/RTD sockets and 45V for the mA/V sockets 
proportional to ground. Thermostats must not be 
connected to any other voltage sources during test. 
About insertion tubes, insulation plugs, well and 
sensor: 
 Never leave hot insertion tubes which have been 
removed from the calibrator unsupervised – they may 
constitute a fire hazard. 
  If you intend to store the calibrator in the optional 
 aluminium carrying case after use, you must ensure 
 that the instrument has cooled down to a temperature 
 below 100°C/212°F before placing it in the carrying 
 case. 
About the fuses: 
 The fuse box must not be removed from the power 
control switch until the mains cable has been 
disconnected. 
 The two main fuses must be identical and correspond 
to the chosen voltage. 
About the liquid bath: 
 For liquid bath ensure that the sensor is absolutely 
clean and dry as a few drops of water in the well (liquid 
baths) might cause a steam explosion. 
 Do not pour cold fluid into a hot well – it might cause 
an explosion.