EasyManuals Logo

Ametek MiniFlash FP Vision User Manual

Ametek MiniFlash FP Vision
69 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #33 background imageLoading...
Page #33 background image
27
FP(H) VISION
OPERATION MANUAL
© AMETEK, Inc.
If the flashpoint is not known enter 10°C for the start temperature for the first time. The result should be taken as an
approximate one. Perform a test again with a start temperature 18 °C (64,4 °F) below the flashpoint. If the start
temperature is set too high, above the flashpoint of the sample, the true flashpoint can’t be detected since it is
somewhere below the adjusted start temperature.
If the start temperature is set to low, the test time is unnecessarily long. For a quick estimate of the flashpoint of a
sample the faster screening method can be used to evaluate the range where the flashpoint is present to get an
adequate measuring range.
Tf is the final heating temperature. Tf must be above the expected flashpoint of the sample, otherwise the device will
finish the measurement with the result “no flash”.
Ti-T: This parameter cools the oven below the set temperature before starting the measurement cycle. The possible
settings are 0-9°C.
Schematic heating diagram with the description of the adjustable parameters
Every different adjustment for different samples can be saved as a new template on the unit with different names.
This should allow the user to create own templates which can be overwritten and deleted from the unit.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Ametek MiniFlash FP Vision and is the answer not in the manual?

Ametek MiniFlash FP Vision Specifications

General IconGeneral
BrandAmetek
ModelMiniFlash FP Vision
CategoryTest Equipment
LanguageEnglish

Related product manuals