Table 202 – Guidance and Manufacturer’s Declaration
Electromagnetic Immunity
All Medical Electrical Equipment and Medical Electrical Systems
Guidance and Manufacturer’s Declaration – Electromagnetic Immunity
-
tomer or user of the Reichert 7CR should ensure that it is used in such an environment.
Immunity
Test
IEC 60601
Test Level
Compliance
Level
Electromagnetic
Environment - Guidance
Electrostatic
Discharge
IEC 61000-4-2
Floors should be wood, concrete or ce-
Electrical Fast
IEC 61000-4-4
Lines
100kHz repetition
frequency
Lines
100kHz repetition
frequency
Mains power quality should be that of a
typical residential, commercial or hospital
environment.
Surges
IEC 61000-4-5
Line-to-line
Line-to-ground
Common Mode
Mains power quality should be that of a
typical residential, commercial or hospital
environment.
IEC 61000-4-11
45°, 90°, 135°, 180°,
225°, 270°, 315°
45°, 90°, 135°, 180°,
225°, 270°, 315°
Mains power quality should be that of a
typical residential, commercial or hospital
environment. If the user of the Reichert
7CR requires continued operation
during power mains interruptions, it is
recommended that the Reichert 7CR be
powered from an uninterruptible power
supply or battery.
Single phase: at 0°
Single phase: at 0°
Interruptions
IEC 61000-4-11
cycles
cycles
Power Frequency
50/60Hz
Magnetic Field
IEC 61000-4-8
30A/m
50 Hz or 60 Hz
30A/m
50 Hz or 60 Hz
be that of a typical residential, commer-
cial or hospital environment.