16060-101 Rev. M
43
Table 204 – Guidance and Manufacturer’s Declaration
Electromagnetic Immunity
Medical Electrical Equipment and Medical Electrical Systems that are NOT Life-supporting
Guidance and Manufacturer’s Declaration – Electromagnetic Immunity
customer or user of the Reichert 7CR should ensure that it is used in such an environment.
Immunity
Test
IEC 60601
Test Level
Compliance
Level
Electromagnetic
Environment - Guidance
Conducted distur-
bances induced
IEC 61000-4-6
150 kHz to 80 MHz
in ISM bands between
150 kHz and 80 MHz
150 kHz to 80 MHz
bands between 150
kHz and 80 MHz
Portable and mobile RF communications equip-
ment should be no closer to any part of the
Reichert 7CR, including cables, than the recom-
mended separation distance calculated from
the equation applicable to the frequency of the
transmitter.
Recommended Separation Distance:
80 to 800 MHz
800 MHz to 2.7 GHz
Where P is the max output power rating of the
-
mitter manufacturer and d is the recommended
-
mined by an electromagnetic site survey, should
be less than the compliance levels in each
frequency range.
Interference may occur in the vicinity of equip-
ment marked with the following symbol.
Radiated RF
Electromagnetic
Fields
IEC 61000-4-3
80 MHz to 2.7 GHz
80 MHz to 2.7 GHz
80 MHz to 2.7 GHz
80 MHz to 2.7 GHz
Note 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
Note 2:
should be observed to verify normal operation. If abnormal performance is observed, additional measures many be
necessary, such as re-orienting or relocating the ME Equipment or ME System.
13,553 MHz to 13,567 MHz; 26,957 MHz to 27,283 MHz; and 40,66 MHz to 40,70 MHz. The amateur radio bands
between 0,15 MHz and 80 MHz are 1,8 MHz to 2,0 MHz, 3,5 MHz to 4,0 MHz, 5,3 MHz to 5,4 MHz, 7 MHz to 7,3
MHz, 10,1 MHz to 10,15 MHz, 14 MHz to 14,2 MHz, 18,07 MHz to 18,17 MHz, 21,0 MHz to 21,4 MHz, 24,89 MHz
to 24,99 MHz, 28,0 MHz, to 29,7 MHz and 50,0 MHz to 54,0 MHz.