Guidance Tables (continued)
Table 202 – Guidance and Manufacturers Declaration
Electromagnetic Immunity
All Equipment and Systems
Guidance and Manufacturers Declaration – Electromagnetic Immunity
The Phoroptor VRx is suitable for use in electromagnetic environment specied below. The
customer or user of the Phoroptor VRx should ensure that it is used in such an environment.
Immunity
Test
IEC 60601
Test Level
Compliance
Level
Electromagnetic
Environment
- Guidance
ESD
IEC 61000-4-2
±6kV Contact
±8kV Air
±6kV Contact
±8kV Air
Floors should be wood, concrete, or ceramic
tile. If oors are synthetic, the R/H should be at
least 30%.
EFT
IEC 61000-4-4
±2kV Mains
±1kV I/Os
±2kV Mains
±1kV I/Os
Mains power quality should be that of a
typical residential, commercial, or hospital
environment.
Surge
IEC 61000-4-5
±1kV Differential
±2kV Common
±1kV Differential
±2kV Common
Mains power quality should be that of a
typical residential, commercial, or hospital
environment.
Voltage
Dips/Dropout
IEC 61000-4-11
>95% Dip
for 0.5 Cycle
60% Dip
for 5 Cycles
30% Dip
for 25 Cycles
>95% Dip
for 5 Seconds
>95% Dip
for 0.5 Cycle
60% Dip
for 5 Cycles
30% Dip
for 25 Cycles
>95% Dip
for 5 Seconds
Mains power quality should be that of a
typical residential, commercial, or hospital
environment. If the user of the Phoroptor
VRx requires continued operation during
power mains interruptions, it is recommended
that the Phoroptor VRx be powered from an
uninterruptible power supply or battery.
Power Frequency
50/60Hz
Magnetic Field
IEC 61000-4-8
3A/m 3A/m
Power frequency magnetic elds should be that
of a typical residential, commercial, or hospital
environment.
107
16241-101 Rev. D