Adjust Values
Now that you aligned the Phoroptor Head and imported data, start the refraction.
Adjusting a Refraction From Previously Saved, Transferred, or Input Data
1. Touch the drop-down list in the main display area or press the AR or LM button on the keypad to
select the data you want to use to start the refraction (e.g., LM, AR).
Note: The data in the boxes is transferred to the refraction data elds above and become
active.
Note: If the Subjective data set was previously modied, the AR or LM data is not copied to it.
Note: The data is labeled the same way as in the data storage area (e.g., LM).
Note: The Phoroptor Head automatically switches in the lenses to reect the information in the
active eld.
2. Adjust measurements from this starting point.
3. Once you make an adjustment, the label changes from the previous label (e.g., LM, AR) to
Subjective.
Figure IU-12, Values Table
Eye Selection
1. Press the R, L, or B (Binocular) button to open the apertures for the right eye, left eye, or both
eyes, respectively.
Note: The aperture for the inactive eye automatically closes during refraction.
Note: The default setting for the open aperture of a new exam is selected in the Setup Menu
after turning on the unit.
Note: The Refraction Sequence is determined by what you selected in the Setup Menu.
Instructions for Use (continued)
59
16241-101 Rev. D