Cross Cylinder - Smart Test
Axis Adjustment
1. Touch XCyl on the menu bar at the bottom of the main screen.
2. Touch SMART to begin the Smart Cylinder test.
The Smart Cross Cylinder test is performed monocularly.
Spherical equivalent is maintained during the Cylinder power adjustment.
If no Cylinder power is present, the Astigmatism check is done rst, followed by the Axis
adjustment.
If default charts are selected for the XCyl test, these charts are presented to the patient.
Otherwise, select a chart from the available charts section.
3. Turn the CONTROL KNOB to change the position of the conventional Cross Cylinder.
4. Ask the patient: “Which is better (sharper and clearer), 1 or 2?”
5. Change the Cylinder Axis in the direction the patient prefers.
6. Press the CONTROL KNOB, or touch the NUMBER (1 or 2), to enter the patient’s preferred
selection.
Note: Touch Test to switch the type of XCyl test (Manual/Smart/Split Cyl).
Figure IU-20, Cross Cylinder - Smart Test Screen