Data Output - Printing
1. Connect an MCP1000 printer to the system. Refer to Figure IS-06 for the connection port.
2. Select the Device driver in the Ports Settings panel and touch APPLY.
Note:
If using a wireless connection, congure the port for a wireless setup and pair a Bluetooth
adapter. Refer to Figure IS-14.
Figure IU-46, Printer Settings Screen
3. Select the Printer options in the Setup Menu. Refer to Figure IS-24.
4. Select the data you want to print in the lower right corner of the screen. The data printing
options include: (Please refer to Figure IS-24C.)
Subj, AR, and LM values.
Visual Acuity, FCC, NRA/PRA, Accommodation, Phoria, and Vergence values.
The Print Format drop-down list options include:
Far - Prints values for Far.
Near - Prints values for Near.
Far/Near - Prints values for Far and Near.
Far/Add - Prints values for Far and Near Vision Addition.
5. Press the OUT button, or touch OUT on the main screen, to immediately send the data to the
printer when the exam is complete.
Instructions for Use (continued)
96 16241-101 Rev. D