Near Vision Test (continued)
Fused Cross Cylinder Test (continued)
6. Touch SET ADD to set the patient’s Add value to that determined by the Fused Cross Cylinder
test.
Figure IU-23, Fused XCyl Test
Instructions for Use (continued)
76 16241-101 Rev. D