Safety instructions qTOWER
3
touch/qTOWER
3
G touch
12
2.6.2 Safety instructions - Protection against explosion and fire
The qTOWER³ touch must not be operated in an explosive environment.
The operating personnel must be familiar with the location of the fire-fighting equip-
ment in the operating room of the device.
The qTOWER³ touch must not be operated with flammable, explosive or volatile sub-
stances.
2.6.3 Safety instructions - electrical equipment
Work on electrical components of the qTOWER³ touch may only be carried out by a
qualified electrician in accordance with the applicable electrical engineering rules. Life-
threatening electrical voltages may occur in the interior of the qTOWER³ touch!
Observe the following:
Any work on the interior of the device may only be carried out by the customer
service of Analytik Jena AG and specially authorized technicians.
The electrical components must be checked regularly by a qualified electrician. Any
defects, such as loose connections, faulty or damaged cables, must be repaired
without delay.
Do not replace the removable mains cable of the device by a mains cable that does
not meet the specifications (without protective ground conductor).
Before opening the device it must be switched off from the power switch and the
mains connector must be disconnected from the mains outlet!
The qTOWER³ touch must be switched off immediately using the power switch and
the power supply disconnected from the mains if there is any interference with the
electric components.
The qTOWER³ touch fulfills the corresponding requirements for interference emis-
sion and resistance from the IEC 61326 series. Assess the electromagnetic envi-
ronment of the device before operation.
Do not operate the device near sources of strong electromagnetic radiation (e.g.
unshielded, deliberately operated high frequency source) as these may affect prop-
er operation of the device.

Table of Contents