EasyManua.ls Logo

Analytik Jena qTOWER3 touch - Connections and control elements

Default Icon
41 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Commissioning qTOWER
3
touch/qTOWER
3
G touch
18
Energy supply
Operating voltage
Line frequency
100 V, 115 V, 230 V
50 60 Hz
Power consumption max. 850 W
Mains voltage
Instrument fuses
100 V, 115 V, 230 V 2 x 10 AT / 250 V
4.1.3 Space requirement
NOTICE
When opening the qTOWER³ touch the upper device hood tilts back. Allow adequate
space for this.
qTOWER
3
touch Height [cm] Width [cm] Depth [cm]
closed 59 27.5 27.5
open 70 27.5 50
The open device has a space requirement of 70 cm x 27.5 cm x 50 cm (H x W x D).
Allow adequate space for placement of the device. In addition, keep a safety distance
of at least 10 cm to other equipment or walls!
4.2 Connections and control elements
1
Upper part with fluorescence-
spectrometer
2
Integrated tablet PC
3
Lock with handle
4
LED for status display
5
Lower part with thermal cycler
Fig. 3 qTOWER³ touch front view
Switching-on conditions
Instrument fuses

Table of Contents

Related product manuals