Andor Technology Zyla sCMOS 4.2 PLUS - 2.6.2.6 Rolling Shutter External Exposure Triggering (Non-Overlap Mode); Table 16: Timing Parameters

Andor Technology Zyla sCMOS 4.2 PLUS
78 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
version 2.0 rev 18 Feb 2021
36
2.6.2.6 Rolling Shutter External Exposure Triggering (Non-Overlap Mode)
While waiting on the trigger event, the sensor held in a global clear state and is put into a “pre-scan read out
cycle. On detection of the trigger event, Global Clear goes LOW and a frame read out is initiated. This frame
isdiscardedasitdoesnotcontainthecorrectexposureperiod.Readingoutaframeeectivelybeginsanew
exposure. When the external trigger input goes LOW, a signal frame read out phase begins. When the frame
has been read out completely, the Global Clear is held HIGH until the next trigger event is detected.
Theexternaltriggerpulsewidthdenestheexposuretimeforallrows,butisonlycoincidentwiththeexposure
time for Row 1. The exposure for Row 2 will be delayed by one row time relative to Row 1 and so forth.
Theperiodoftheexternaltriggerpulsedenestheoverallcycletime.Ifthewidthofthetriggereventis
less than 3 Rows, the falling edge will be missed and a subsequent falling edge will be required to end the
exposure.
Figure20:Rolling Shutter External Exposure Triggering (Non Overlap Mode)
Table 16: Rolling Shutter External Exposure Triggering (Non Overlap Mode) Timing Parameters
Parameter Minimum Maximum
Exposure 3 Rows 30 s
Cycle Time (EXT Trig Period) Exposure + 1 Frame + 1 Row -
External Start Delay 0 1 Row
EXT Trig Pulse Width 3 Rows 30 s