version 2.0 rev 18 Feb 2021
45
Iftheexposuretimeisgreaterthanaframereadouttime,theexposurestartsrstbypulsingtheGlobalClear.
The reference frame is read out during the exposure such that the end of the reference read out is coincident
with the end of the exposure. The Charge Transfer pulse then goes HIGH to transfer charge from all pixels in the
sensor. Finally, the signal frame is then read as shown below.
Figure 27: GlobalShutterInternalTriggering–LongExposures
Table 24: Global Shutter Internal Triggering - Long Exposures Timing Parameters
Parameter Minimum Maximum
Exposure 1 Frame + 4 Rows 30 s
Cycle Time (1/Frame Rate)
Exposure + 1 Frames + 1 InterFrame + 6
Rows
20,000 s
Acquisition Start Delay 0 1 Row

Table of Contents