BAUTS Bat tery Auto Turn OFF SYS TEM
Syn tax: BAUTS <state>
state: ENABLE or DISABLE
Re marks: En able/dis able the bat tery auto power shut off.
NOTE: Al though GPIB is not avail able un der bat tery op era tion, the
state of this pa rame ter can be changed for later use.
Query: BAUTS?
Re turned
String: BAUTS <state>
BAUTT Bat tery Auto shut off af ter x min utes
SYS TEM
Syn tax: BAUTT <val>
val: 10 to 240 min utes
Re marks:
Au to matically turns the unit off af ter x min utes when op er at ing on bat -
tery power.
NOTE: Al though GPIB is not avail able un der bat tery op er a tion, the
state of this pa ram e ter can be changed for later use.
Query: BAUTT?
Re turned
String: BAUT <val>
BUFF GPIB re sponse buff er ing en abled
SYSTEM
Syn tax: BUFF <s>
s: ON
OFF
Re marks: If BUFF is ON: In the ML243X na tive mode, 488.2 GPIB op er a tion,
when a re quest for data is made the re sponse is put in an out put buffer
ready to be read by the con trol ler. If an other data re quest is made and
the pre vi ous data has not been read out of the out put buffer; the new
data is queued af ter the orig i nal re quest. In this mode of op er a tion the
GPIB re sponse buff er ing en able is ON, and fol low ing the 488.2 spec i fi ca -
tions when ever a re quest for data is made the re sponse should be read.
If BUFF is OFF: In this mode when ever a re quest for data is made, (ex -
cept by se rial poll) the out put buffer is cleared and the only data in the
out put queue will be the re sponse to the last data re quest made. The
6-20 ML2400A OM
ML24XXA NATIVE COMMANDS GPIB OPERATION