tered with this com mand will be avail able for use by the DSP, but will
NOT be saved to the sen sor un til a save com mand (
CFUSAV ) is ex e -
cuted. If the sen sor is changed or power is lost be fore sav ing, all changes
made since the last
CFUSAV will be lost.
The user must en sure that the max i mum num ber of cal fac tor data pairs
en tered into a ta ble is not ex ceeded. Sen sors with a max i mum fre quency
of up to 40 GHz will hold 90 pairs, while sen sors with a max i mum fre -
quency of 50 GHz will hold 110 pairs.
Re lated
Com mands:
CFUSAV
CFUCT Clear cal fac tor table
SEN SOR
Syn tax: CFUCT <s>,<ta ble number>
s: A or B
ta ble
number: 1 to number of ta bles sup ported by the sen sor type
Re marks: Clears the cal fac tor ta ble to one en try for 50 MHz at 100%, but does not
clear the iden tity of the ta ble. The cleared ta ble is au to mat i cally saved to
the sen sor.
CFUID Cal fac tor ta ble iden tity update
SEN SOR
Syn tax: CFUID <s>,<ta ble number>,<identity>
s: A or B
ta ble
number: 1 to number of ta bles sup ported by the sen sor type
identity: Seven characters or until a message ter mi na tor will be read as the
identity.
Re marks: Up dates the seven char ac ter iden tity string. This only af fects the copy
of the cal fac tor ta ble stored in the mem ory of the power me ter. To take
af fect and not be lost, the ta ble must be saved to the sen sor us ing the
CFUSAV com mand.
Query: CFUID? <s>,<ta ble num ber>
Re turned
String: CFUID <s>,<ta ble num ber>,<iden tity>
CFUNITS Cal fac tor dis play units
SEN SOR
Syn tax: CFUNITS <s>,<units>
6-24 ML2400A OM
ML24XXA NATIVE COMMANDS GPIB OPERATION