s: A or B
units: % or PCT
dB or dBm
Re marks: This com mand changes the dis play units of the calfactors be tween ei ther
dB or pecentage. Note that this will also set the form the data is out put
over the GPIB (or RS232) when re quested.
Query: CFUNITS? <s>
Re turned
String: CFUNITS <s>,<units>
CFULD Cal fac tor ta ble bi nary load
SEN SOR
Syn tax: CFULD <s>,<ta ble number>,<length>,<bi nary data>
s: A or B
ta ble
number: 1 to number of ta bles sup ported by the sen sor type
length: Length of mes sage in bytes
bi nary
data: Same data as that re cieved by CFURD
Re marks: Loads bi nary data into the cal fac tor ta ble. This command will
automatically save the data to the sensor.
CFUPT Preset cal fac tor table
SEN SOR
Syn tax: CFUPT <s>,<ta ble number>
s: A or B
ta ble
num ber: 1 to number of ta bles sup ported by the sen sor type
Re marks: Presets the cal fac tor ta ble to the fac tory set tings. The pre set ta ble is
auto mati cally saved to the sen sor.
CFURD Cal fac tor ta ble bi nary read
SEN SOR
Syn tax: CFURD <s>,<ta ble number>
s: A or B
ML2400A OM 6-25
GPIB OPERATION ML24XXA NATIVE COMMANDS