EasyManua.ls Logo

Anritsu ML2400A Series - Page 122

Anritsu ML2400A Series
275 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
DTRGD Dis play Trig ger De lay PRO FILE SETUP
Syn tax: DTRGD <val>
val: 0.0 to 7.0 sec onds
Re marks: The de lay time from the trig ger point to when the pro file starts to be
drawn (re fer to Fig ure 4-4, page 4-20 ).
Ex am ple: DTRGD 1.25MS Sets the dis play trig ger de lay to 1.25 ms.
Query: DTRGD?
Re turned
String: DTRGD <val>
DUTY Duty cycle
SENSOR
Syn tax: DUTY <s>,<duty_cy cle>
s: A or B
duty cy cle: 0.1 to 100%
Re marks: Ap plies a duty cy cle to the se lected sen sor. An off set will be ap plied
based on the en tered value.
Ex am ple: DUTY A,50
Speci fies a duty cy cle of 50% that will al ter the dis played read ings by ap -
proxi mately +3.01 dB.
Re lated
Com mands: DUTYS
Query: DUTY? <s>
Re turned
String: DUTY <s>,<duty cy cle>
DUTYS Duty cycle state
SENSOR
Syn tax: DUTYS <s>,<state>
s: A or B
state: ON or OFF
Re marks: Turns on or off the duty cy cle for the se lected sen sor.
Re lated
Com mands: DUTY
ML2400A OM 6-35
GPIB OPERATION ML24XXA NATIVE COMMANDS

Table of Contents