EasyManua.ls Logo

Anritsu ML2400A Series - Page 128

Anritsu ML2400A Series
275 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Re marks: When pro file DATA HOLD mode is set to Min/Max points, this com -
mand is used to re set the min/max val ues for each data point.
Re lated
Com mands: GRPIX
GRAUTO Auto scal ing
SYS TEM
Syn tax: GRAUTO
Re marks: Auto scale for all graphic modes (Pro file, Source Sweep and Power vs.
Time). This com mand auto scales the y axis only based on the cur rently
dis played data.
GRAVG Av er age be tween pro file cur sors
PRO FILE SETUP
Syn tax: GRAVG <state>
state: ON or OFF
Re marks: Turns on or off av er ag ing be tween cur sors. The data re turned by the
GRDRQ com mand in cludes the av er age of all data points be tween the
cur sors if GRAVG is turned ON.
Re lated
Com mands: GRDRQ
Query: GRAVG?
Re turned
String: GRAVG <state>
GRCP Con nect points on pro file
PRO FILE SETUP
Syn tax: GRCP <state>
state: ON or OFF
Re marks: When set to ON, cre ates a line graph by con nect ing the pro file data
points to gether. The de fault is ON.
Query: GRCP?
Re turned
String: GRCP <state>
ML2400A OM 6-41
GPIB OPERATION ML24XXA NATIVE COMMANDS

Table of Contents