LINK Trig ger linking TRIGGER
Syn tax: LINK <state>
state: ON or OFF
Re marks: This will link the trig ger set-up be tween Pro file mode and Read out
mode so that the sam ple de lay and the gate width will agree. A change to
the trig ger set-up in ei ther Read out or Pro file sys tem set-up will af fect
ei ther dis play mode.
Query: LINK?
Re turned
String: LINK <state>
LLIM Set Low lim its
CHAN NEL
Syn tax: LLIM <c>,<val>
c: 1 or 2
val:
Re marks: Sets the low limit. The LLIMS com mand turns the lim its on and off. It is
not nec es sary to en ter the units as the limit value is checked against the
dis played value.
There fore, if the lim its have been set for –10 dBm ( LLIM 1, –10) and the
dis play units are sub se quently changed from dBm to Watts, the sys tem
still checks for the read ing to rise above –10, even though the dis play
units type has been changed.
Re lated
Com mands: LLIMS
Query: LLIM? <c>
Re turned
String: LLIM <c>,<val>
LLIMS Turn on/off low lim its
CHAN NEL
Syn tax: LLIMS <c>,<state>
ML2400A OM 6-53
GPIB OPERATION ML24XXA NATIVE COMMANDS
Units Min Max
dBm –99.99 +99.99
dBmV –53.00 147.00
dB
mV 7.00 207.00
Watts 0.0 50.0