Re marks: Avail able in graph modes only. Out puts in ASCII form the min graph
data. The for mat is as fol lows:
MNGD <num ber_of_el e ments>,<el e ment_1>,<el e ment_2>,<el e -
ment_n>...<\n>
The first num ber in the string is the num ber of el e ments to fol low, and
is al ways 200 for the ML2400A Se ries.
MNMXS Track min and max val ues
CHAN NEL
Syn tax: MNMXS <c>,<state>
c: 1 or 2
state: ON or OFF
Re marks: Turns ON or OFF the min/max track ing for the speci fied chan nel. The
MMRST com mand re sets the val ues.
Re lated
Com mands: MMRST
Query: MNMXS? <c>
Re turned
String: MNMXS <c>,<state>
MODDEL Mo dem re dial de lay time
SYSTEM
Syn tax: MODDEL <value>
value: 1 to 10
Re marks: Sets the autodial de lay be tween retrys. The value is the num ber of min -
utes to de lay be tween each autodial re try af ter a fail ure to con nect. This
in ter val can be set from 1 to 10 min utes. See Sec tion 5-10 for more in -
for ma tion on mo dem op er a tion.
Query: MODDEL?
Re turned
String: MODDEL <value>
MODINIT Ini tial ize mo dem
SYS TEM
Syn tax: MODINIT
Re marks: Ini tial izes the mo dem con nected to the ML2400A se rial port. See Sec tion
5-10 for more in for ma tion on mo dem op er a tion.
ML2400A OM 6-55
GPIB OPERATION ML24XXA NATIVE COMMANDS