Out put Re quests There are three com mands to re quest out put from the ML4803A: OPW for a
read ing, ODT for the cal fac tor, off set and ref er ence val ues, and OMR for mem -
ory store set tings. If these out put re quests are re ceived si mul ta neously, only
the data for the com mand re ceived last will be avail able.
Un sup ported
Com mands
The fol low ing ML4803A com mands are not sup ported in the ML2400A Se ries
Power Me ter GPIB in ter face:
PCT
VSW
RDB
DBV50
DBV75
VLT50
VLT75
These com mands are read in without er rors, but are ig nored by the sys tem.
AVE Sen sor av er ag ing set ting.
Syn tax: AVE<num ber>
num ber: 0 = Av er aging OFF
9 = HOLD. Holds the pres ent av er aged read ing.
1 = Av er age for 1 sec ond (ML2400A Re peat av er age num ber of 25).
2 = Av er age for 2 sec onds (ML2400A Re peat av er age num ber of 70).
3 = Av er age for 5 sec onds (ML2400A Re peat av er age num ber of 128).
4 = Av er age for 10 sec onds (ML2400A Re peat av er age num ber of 256).
Re marks: The ML4803A averages for a pe riod of time. The ML2400A sets the av -
er ag ing to re peat av er ag ing for a num ber of read ings.
CAL Set the user cal fac tor value.
Syn tax: CAL<value>
value: Cal fac tor value in dB