When se lect ing GPIB em u la tion modes, the in stru ment configures it self
to the pre set con di tions of the in stru ment to be em u lated. For ex am ple,
when se lect ing HP 438A em u la tion, the front panel menus pass through
the pre sets for the HP 437B (which pre sets sen sor A to dBm) then se -
lects HP 438A em u la tion (which pre sets sen sor A to Watts).
NOTE
This com mand must be en tered us ing the 488.2 for mat;
that is, EMUL<ws><mode> (<ws> = white space).
MCA Set the cal fac tor value at the spec i fied mem ory lo ca tion
in dBm.
Syn tax: MCA<mem><;><value>
mem: Mem ory lo ca tion 1 to 30.
value: Cal fac tor value in dBm.
Re marks: Set the cal fac tor value at mem ory store ad dress <mem> to <value>
dBm.
MCC Clears the cal fac tor value at the spec i fied mem ory lo ca -
tion.
Syn tax: MCC<mem>
mem: Mem ory lo ca tion 1 to 30.
Re marks: Clears the cal fac tor value at mem ory store <mem> to 0.0 dBm.
MCO Clears the off set value at the spec i fied mem ory lo ca tion.
Syn tax: MCO<mem>
mem: Mem ory lo ca tion 1 to 30.
Re marks: Clears the off set value at mem ory store <mem> to 0.0 dBm.
MCQ Clears the fre quency value at the spec i fied mem ory lo ca -
tion.
Syn tax: MCQ<mem>
mem: Mem ory lo ca tion 1 to 30.
Re marks: Clears the fre quency value at mem ory store <mem> to 0.1MHz.
6-90 ML2400A OM
ML4803A EMULATION COMMANDS GPIB OPERATION