Sprint Pro Set Up & Menu Configuration Options
CO
2
Zero
In normal operation Sprint Pro will perform its zeroing function whilst in clean
air.
However if it is not possible to undertake the zeroing function in clean air then
a CO
2
scrubber can be utilized.
In clean air: Sprint Pro will undertake the zeroing function in clean air and a CO
2
scrubber must not be used whilst zeroing.
With scrubber: Sprint Pro will undertake the zeroing function in clean air and a
CO
2
scrubber must be used whilst zeroing.