EasyManuals Logo

Anton Sprint Pro User Manual

Anton Sprint Pro
44 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #15 background imageLoading...
Page #15 background image
11
Sprint Pro Set Up & Menu Configuration Options
CO
2
Zero
In normal operation Sprint Pro will perform its zeroing function whilst in clean
air.
However if it is not possible to undertake the zeroing function in clean air then
a CO
2
scrubber can be utilized.
In clean air: Sprint Pro will undertake the zeroing function in clean air and a CO
2
scrubber must not be used whilst zeroing.
With scrubber: Sprint Pro will undertake the zeroing function in clean air and a
CO
2
scrubber must be used whilst zeroing.