20
Performing tests Sprint Pro
Place the probe in the area of inspection for several seconds before moving it to
other locations. Sprint Pro will emit continuous clicks like a Geiger counter. If
higher gas levels are detected the bar graph will increase in readings and the
sounder will increase in pitch.
Press ESC to quit the test.
As with many types of instrumentation the GEP may be
damaged by impact. If the probe is dropped or suffers
significant impact in another way check its operation by
first plugging it into the Sprint Pro and confirming that it
is recognised by the analyser.
If the Sprint Pro detects a fault with the GEP it will display
a warning advising that the GEP is faulty and the test will
be aborted.
If the GEP was subjected to a large impact, does not pass the described test
above or is physically damaged please return to a qualified service centre for
repair/replacement.
8. Ambient air monitor (not available on Sprint Pro 1)
Note: CO
2
readings detailed in this section will only be displayed where a CO
2
sensor
is fitted.
1. Select Ambient Air Monitor from the Test Menu.
2. If a ‘zero’ has been performed since switch on, the pump will switch
on in readiness for test.
If a ‘zero’ has not been performed since switch on, Sprint Pro will
initiate the zeroing process. Press the
key to confirm you are in
clean air and if successful the pump will switch on in readiness for test.
Note: The sound of the pump operating does not indicate the test has begun.
3. Connect probe to the Sprint Pro.
4. Use the soft keys
and to select duration required, the interval for
sampling will be displayed on the screen. The duration can be adjusted
with the soft keys, from a minimum of 15 minutes (sampling every
minute) to a maximum of 7 days (sampling every 30 minutes).
Every sample consists of the peak and average over the sample period.
5. Press
to proceed with the test.
6. Then press soft key
to confirm Sprint Pro has sufficient battery life.
Note: Ensure the Sprint Pro has sufficient battery life for the duration of the
test or connect the charger.
Gas level
Gas Escape
14:49:45 23-Jul-2018
LOW
HIGH
MED

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Anton Sprint Pro and is the answer not in the manual?

Anton Sprint Pro Specifications

General IconGeneral
BrandAnton
ModelSprint Pro
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals