Appendix A
Replacing the Halogen Lamp
Applied Biosystems 7300/7500/7500 Fast Real-Time PCR System Installation and Maintenance Guide 119
Notes
A
2. Open the access door to the 7300/7500/7500
Fast system.
a. Insert a thin screwdriver into the keyhole on
the edge of the
access door,
then push to
unlatch the door.
b. Open the
access door
.
PHYSICAL INJURY
HAZARD. The 7300/7500/7500 Fast system
and lamp are hot! The lamp can become very hot
while in use. Allow sufficient time for the lamp
to cool, and put on protective, powder-free
gloves before handling it.
3. Remove the lamp from the instrument:
a. Slide the lamp release lever downward.
b. Firmly grasp the lamp and lift it up and out
of the slotted mount.
Do not touch the lamp
without powder-free gloves. Finger prints
shorten the lamp life.
4. Inspect the lamp for signs of failure (carbon
typically coats the inside of failed lamps).
5. Place the new lamp into the instrument:
a. Slide the lamp release lever upward.
b. Firmly grasp the lamp, place it into the
slotted mount, then carefully slide the lamp
downward into place.
a b
Good bulb
Failed bulb
Replace
a b

Table of Contents

Other manuals for Applied Biosystems 7500 Fast

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Applied Biosystems 7500 Fast and is the answer not in the manual?

Applied Biosystems 7500 Fast Specifications

General IconGeneral
BrandApplied Biosystems
Model7500 Fast
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals