Why aren't the tilt function buttons working on my Arjo Medical Equipment hand control?
- BBilly JohnsonSep 6, 2025
If the tilt function buttons on your Arjo Medical Equipment hand control aren't working, verify that the Maxi Sky 2 Plus is in Dual Mode. The maximum allowed height difference between spreader bars may have been reached, causing the sit or recline function to stop. The software may have calculated a weight transfer shift between the ceiling lifts, causing one to become overloaded and stopping the motion. This is a normal safety feature, indicated by the "Overweight" icon on the display. With only the spreader bar connected to the Maxi Sky 2 Plus (no patient) in Dual Mode, press the UP button to raise the spreader bar to its highest position. This will reset the height difference calculation.