001-31249-EN Rev 3 • 08/2019
INSTRUCTIONS FOR USE
Maxi Sky 2 PLUS

Table of Contents

Other manuals for Arjo Maxi Sky 2 PLUS

Questions and Answers

  • D
    Dawn WongSep 6, 2025
    Why aren't the tilt function buttons working on my Arjo Medical Equipment hand control?
    • B
      Billy JohnsonSep 6, 2025
      If the tilt function buttons on your Arjo Medical Equipment hand control aren't working, verify that the Maxi Sky 2 Plus is in Dual Mode. The maximum allowed height difference between spreader bars may have been reached, causing the sit or recline function to stop. The software may have calculated a weight transfer shift between the ceiling lifts, causing one to become overloaded and stopping the motion. This is a normal safety feature, indicated by the "Overweight" icon on the display. With only the spreader bar connected to the Maxi Sky 2 Plus (no patient) in Dual Mode, press the UP button to raise the spreader bar to its highest position. This will reset the height difference calculation.
  • N
    Nichole JamesSep 4, 2025
    What does it mean when the symbol is flashing on my Arjo Maxi Sky 2 PLUS display?
    • J
      Jessica KnightSep 4, 2025
      If the symbol is flashing on the display of your Arjo Medical Equipment, it indicates that a motor duty cycle has been reached. The function that exceeded the duty cycle is disabled, but all other functions remain active. Lower the patient and wait until the symbol disappears before attempting another transfer.
  • C
    Chad GreenSep 2, 2025
    What to do if Arjo Medical Equipment stops during transfer with good battery?
    • D
      Dawn LinSep 2, 2025
      If your Arjo Medical Equipment ceiling lift stops during patient transfer despite showing a good battery charge, contact your local Arjo agent to have the batteries replaced.
  • R
    richardphelpsAug 28, 2025
    What to do if the UP button on the Arjo Maxi Sky 2 PLUS Medical Equipment hand control doesn't work?
    • D
      Daniel KellerAug 28, 2025
      If the UP button on your Arjo Medical Equipment hand control isn't working, check that the straps are completely wound inside the ceiling lifts, as the upwards motion may have reached its highest point. Ensure the straps aren't twisted, which could activate the "end of stroke" limiter. If the display shows a malfunction code, contact your local Arjo agent. Verify the Maxi Sky 2 Plus is in "Dual Mode". Check if the auxiliary UP button on the ceiling lift (near the display) works; if it does, the hand control may be the issue. Contact your local Arjo agent to have the hand control replaced.
  • R
    Ricky CochranAug 26, 2025
    What to do if the DOWN button on the Arjo Maxi Sky 2 PLUS Medical Equipment hand control doesn't work?
    • W
      websterkevinAug 27, 2025
      If the DOWN button on your Arjo Medical Equipment hand control isn't working, check if the strap is completely unwound from the ceiling lift. If so, the downwards motion has reached its lowest point. Ensure there is more than 1.5 kg (3.3 lb) suspended from the lifting strap, as the ceiling lift requires minimum tension to operate. Verify if the Maxi Sky 2 Plus is in Dual Mode. If the unit display shows a malfunction code, contact your local Arjo agent. Check if the auxiliary DOWN button on the ceiling lift (near the display) works; if it does, the hand control may be faulty. Contact your local Arjo agent to have the hand control replaced.
  • J
    Jessica HaynesAug 25, 2025
    Why do the batteries of my Arjo Maxi Sky 2 PLUS Medical Equipment discharge quickly?
    • P
      Paula AguirreAug 25, 2025
      If the batteries of your Arjo Medical Equipment are always discharged after a few transfers, place the Maxi Sky 2 Plus under the charging station for a complete charging cycle of 8 hours. If the ceiling lift does not detect the charging station, contact your Arjo agent for maintenance.
  • E
    ecarrilloAug 22, 2025
    What to do if the Arjo Maxi Sky 2 PLUS Medical Equipment ceiling lift does not work with the hand control?
    • S
      Stacey OchoaAug 22, 2025
      If the Arjo Medical Equipment ceiling lift isn't responding to the hand control, ensure that the ‘Power On’ switches on both ceiling lifts are pushed upward. Check if the auxiliary UP/DOWN buttons on the ceiling lifts (near the display) are working. If those work, the hand control might be the issue. Contact your local Arjo agent to replace the hand control.

Summary

Safety Instructions

Important Safety Precautions

Critical warnings and precautions for safe operation and handling.

General Instructions

Safe Working Load and Safety Practices

Covers SWL, battery, shock, fire, and environmental safety guidelines.

Hand Control

Control Panel

Preparation

Powering Up and Pre-Use Checks

Steps for powering on the lift and essential checks before first and every use.

How to Use the Maxi Sky 2 Plus

Dual and Single Mode Operation

Differentiates between Dual Mode (bariatric) and Single Mode (non-bariatric) use.

Usage Monitoring and Emergency Safety

Sleep Mode, Usage Data, and Emergency Stop

Instructions on sleep mode, accessing usage data, and using the emergency stop system.

Emergency Lowering and Battery Info

Emergency Lowering Procedure and Brake

Step-by-step guide for safely lowering a patient in an emergency and brake function.

Battery Charging Systems

Operating in Dual Mode

Movement Controls and Patient Tilt

Explains vertical movement, horizontal movement, and patient tilt functions in Dual Mode.

Dual Mode Operation Details

Sling Selection and Attachment

Spreader Bar Sling Attachment

Spreader Bar Description

Sling Attachment Methods

Transferring from Seated to Lying

Seated to Lying Transfer Steps

Detailed procedure for transferring a patient from a seated position to a lying position.

Transferring from Lying to Seated

Lying to Seated Transfer Steps

Detailed procedure for transferring a patient from a lying position to a seated position.

Bed Repositioning

Bed Turning

Limb Support

Lateral Transfers

Switching Between Modes

Mode Switching Procedures

Switching to Single Mode

Procedure to reconfigure the lift from Dual Mode to Single Mode.

Switching to Dual Mode

Reconfiguring to Dual Mode

Procedure to reconfigure the lift from Single Mode to Dual Mode.

Operating in Single Mode

Care and Maintenance

Cleaning the Ceiling Lifts

Instructions for cleaning the lift and its accessories between patient uses.

Disinfection Procedures

Maintenance and Support

Preventive Maintenance Schedule

Troubleshooting

Common Problems and Solutions

Identifies common issues and provides steps to check or resolve them.

Technical Data

Arjo Maxi Sky 2 PLUS Specifications

General IconGeneral
TypeCeiling Lift
Safe Working Load272 kg (600 lbs)
Maximum Load Capacity272 kg (600 lbs)
Lifting Capacity272 kg (600 lbs)
Power Supply24 V DC
Emergency StopYes
Battery BackupYes
Control TypeHand control
Track OptionsStraight, curved tracks
Emergency LoweringYes
Safety FeaturesOverload protection, emergency stop