Asiga MAX - To Calibrate the MAX Internal Radiometer

Asiga MAX
41 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Asiga MAX User Guide
Copyright ASIGA © 2017 VERSION 1.04
To calibrate the MAX internal radiometer:
1. Go to menu item Maintenance > LED > Calibrate Radiometer. Click “Continue” at
the warning screen.
2. Hold the radiometer head over the build aperture so that the sensor is in the build
area.