Asiga MAX User Guide
Copyright ASIGA © 2017 VERSION 1.04
16. Pressing Cancel will stop the “detect zero position” process and skip to the “Home
vertical axis?” screen.
17. After the process is bypassed or completed, click “Yes” to home the vertical axis or
“No” to leave it at the current position.
Zero Position Encoders
The MAX is equipped with four position encoders built into the corners of the glass base-
plate. The position encoders should be zeroed after transporting the machine. The position
encoders may drift so they should also be zeroed from time-to-time by the user. The
procedure takes less than 30 seconds. The procedure for zeroing the position encoders is
as follows:
1. Open the menu item Maintenance > Position Encoders: