46
HI-Limit Current setting
Advance the menu to the HI-Limit parameter. The display will show:
HI-Limit = XXXX M
Range : 1 - 9999
Use the Numeric keys to enter the HI-Limit setting, then press the ENTER key. The
program will store the HI-Limit setting and advance to the LO-Limit parameter
automatically. The unit is M and 1 M per step.
Press the EXIT key to exit from the setting mode to the operation mode if all parameters
have been set.
LO-Limit Current setting
Advance the menu to the LO-Limit parameter. The display will show:
LO-Limit = XXXX M
Range : 1 - 9999
Use the Numeric keys to enter the LO-Limit setting, then press the ENTER key. The
program will store the LO-Limit setting and advance to the Delay Time parameter
automatically. The unit is M and 1 M per step.
Press the EXIT key to exit from the setting mode to the operation mode if all parameters
have been set.
Delay Time setting
Advance the menu to the Delay Time parameter. The display will show:
Delay Time = XXX.X s
Range : 0.5 - 999.9 0 = Constant
Use the Numeric keys to enter the Delay Time setting, then press the ENTER key. The
program will store the Delay Time setting and advance to the Scanner Channel parameter
automatically. The unit is second and 0.1 second per step.
Press the EXIT key to exit from the setting mode to the operation mode if all parameters
have been set.
Delay Time selects the time when the instrument will compare actual readings to the HI-
Limit and LO-limit settings. This delay allows charging current to stabilize before a test
judgment is made. The capacitance of the DUT will dictate what delay setting is required
to perform an accurate IR measurement.