The phase measurement is always the phase of the selected channel,
using another channel as a reference. Pressing the INPUT A button
will measure the phase of input channel A, using input channel B as a
reference. Pressing the INPUT B button will measure the phase of
input channel B, using input channel A as the reference.
Pressing GEN after selecting the channel for measurement will provide
the following measurements:
.
A-G or B-G This option will measure the phase of the selected
input with the same channel gen-mon path as a reference. This
is used to measure phase shift through a device (input to
output).
.
GA or GB This option will measure the phase of the selected
gen-mon channel with the other gen-mon channel as the
reference.
The lower left soft key controls the range over which the phase
measurement is displayed. The allowable ranges are +180 to –180
degrees, -270 to +90 degrees, and –90 to +270 degrees. The latter
two ranges should generally be used for polarity checks, where the
reading will either be very close to 0 or very close to 180 degrees. If
the +180 to -180 degree range is used for these measurements,
readings near zero could read on either end of the range.
The lower right soft key controls the averaging of the readings. If it is
set to ‘NOAVERAGE’, the phase readings are updated at a rate of
approximately three per second. If it is set to ‘AVERAGE’, the display
shows a running average of the last eight readings, to reduce the
effects of digit bobble. The ‘AVERAGE’ mode is particularly useful
when adjusting azimuth in analog tape machines.
In bargraph mode, the AMPLITUDE keys are used to set the endpoints
of the graph. After selecting the endpoint to change by pressing the
appropriate soft key, the ±10 dB keys will change the endpoint in
increments of 90 degrees. The INC and DEC buttons will change the
endpoint in increments of 10 degrees.
To obtain the specified accuracy, analog inputs must have amplitude
greater than about 20 mV and must have signal-to-noise ratio greater
than about 40 dB. If either input drops below about 8 mV (or below
4
Operation
Controlling the Analyzer : Phase Operation
Portable One Plus Access User's Manual 4-41
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Audio Precision Portable One Plus Access and is the answer not in the manual?

Audio Precision Portable One Plus Access Specifications

General IconGeneral
BrandAudio Precision
ModelPortable One Plus Access
CategoryTest Equipment
LanguageEnglish

Related product manuals