USER MANUAL ME-01 35
17.1 Scale calibration (CALIb)
Press MENU key.
Press
T
key when CALIb function appears.
The following options will be displayed:
-CAL on – calibration with external
recommended standard of mass (see technical
data).
-CAL StP – calibration with external weight,
confirmation of succesive steps - MENU key,
out – leave without changes
Press
T
key when CAL StP option appears
(calibration in two steps).
Press
T
key when weight value used for
calibration is indicating or use othEr option and
inscribe proper value ( keys
0
, ,
T
)
Press MENU and wait for writing zero to the
scale.
When LOAD message appears put standard of
mass on the pan. Press MENU key (CAL on
doesn’t need pressing MENU key).
Wait until internal calibration is finished and zero
indication is displayed.