Service Program
3
3 - 2 Perfusor® compact, 2.1 gb
WARNING
NEVER RUN SERVICE MODE WHEN A PATIENT IS CONNECTED!
DO NOT CONNECT THE MFC SERVICE CONNECTOR OR THE SERV-
ICE CABLE WHEN A PATIENT IS CONNECTED TO THE UNIT! FIRST
SWITCH THE UNIT OFF BEFORE ANY FURTHER USE AFTER WORK
-
ING WITH THE MFC SERVICE CONNECTOR. CARRY OUT A CHECK
ACCORDING TO THE PROCEDURAL INSTRUCTIONS FOR INSPEC
-
TION AFTER THE SERVICE PROGRAM WAS RUN (see „Procedural
Instructions for Inspection after Modifications via the Service
Program“ ➨ p. 3 - 11).
When the Service Program is installed and the PC is connected to
the Perfusor compact, the following functions can be executed:
- Drive calibration
- Reading / loading pump data
- Displaying operation values
- Displaying and changing parameters
- Documentation of pump hardware modifications
- Saving all data to a diskette, hard disk or similar
Installation
1. Insert diskette.
2. Start the File Manager or Windows Explorer.
3. Start Setup.exe with a double-click. The directory
C:\PLC_SERV is created automatically by the program. Cali
-
bration and default data is also saved in this directory (if no
other directory is specified). The directory name can be
changed without any problems.
Note
The system configuration of the PC is not changed when the Serv-
ice Program was installed.
Uninstall
1. Delete the Plc_serv.exe file to uninstall.
Note
If the complete directory PLC_SERV is deleted all unit data is de-
leted from the PC.