Perfusor® compact, 2.1 gb 4 - 1
4
4 - Unit Elements
Perfusor® compact
4.1 Fundamental Repair Information
Battery Pack and Batteries
Designation Ord. No.
Battery pack . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3450 1690
Note
Always disconnect unit from mains.
Prior to repair:
1. Switch off the Perfusor® compact.
2. Disconnect the unit from mains.
3. Remove batteries to avoid short circuits or consequential
damage.
Note
The battery may only be removed when the device is switched off
as otherwise alarm 022 is displayed upon startup. Press the ON-/
OFF-button to delete the alarm 022 until the alarm symbol is no
longer displayed. If the alarm 105 is triggered afterwards switch
the unit off.
Before startup:
4. If batteries are used switch the device first on without mains
connection. If the battery pack is used, then the device is to
be switched on with mains connection.
Note
Defective batteries must be disposed of according to the regula-
tions, e.g. return to B. Braun (see „Return of Spare Parts and Test
Equipment“ p. 0 - 9).
Fitting Plastic Screws
In order to avoid damage to the thread:
Turn anti-clockwise (until the thread is found), then turn clock-
wise to fasten (max. 0.5 Nm).