Procedural Instructions on the TSC
8
Perfusor® compact, 2.1 gb 8 - 3
11. Pull syringe holder.
Staff call: red LED in MFC service connector lights up. Drive
stops.
Note
The following staff call modes can be selected in the Service Pro-
gram if the unit is switched off: static, dynamic with and without
alarm.
12. Pull syringe holder.
Staff call: red LED in MFC service connector lights up. Drive
stops.
13. Switch device off.
14. Disconnect unit from mains.
15. Switch unit on in battery mode.
Pressure Cut-Off
1. Set pressure stage 1 (key sequence F 3 C 1 F START).
CAUTION
The limit values of the Checklist for Checks after Repair do not
correspond to the TSC values.
2. Pump continues to deliver at 96.0 ml/h and switches off when
the specified pressure is reached.
3. Enter value in check list.
4. Set pressure stage 2 (key sequence F 3 C 2 F START).
5. Pump continues to deliver at 96.0 ml/h and switches off when
the specified pressure is reached.
6. Enter value in check list.
7. Set pressure stage 3 (key sequence F 3 C 3 F START).
8. Pump continues to deliver at 96.0 ml/h and switches off when
the specified pressure is reached.
9. Enter value in check list.
WARNING
DANGER OF INJURY! DO NOT OPEN TOGGLE AS CURRENT STEP
GAUGE CAN RELEASE SUDDENLY.
10. Release current step gauge (key sequence F 3 0) with service
connector.
Fig.: 8 - 2
OPS 50 ml slot