EasyManua.ls Logo

B. Braun Perfusor compact - Unit Calibration Procedures; General Calibration Setup; Detailed Calibration Steps

B. Braun Perfusor compact
82 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Service Program
3
3 - 14 Perfusor® compact, 2.1 gb
Syringes 50 ml Free Type / Syringe Type Number
1. Press keys 7 C {number of the 50 ml free type} F – syringe is
accepted (if an alarm is triggered, then the syringe selection
setting is not correct!) Check whether the free type was
marked on the unit (please see syringe table).
2. Reset syringe types after the test is finished!
Syringe Selection
Reset syringe selection according to the condition as the unit was
delivered.
Unit Calibration
General
1. Connect unit to PC and start the Service Program (see „Wor-
king with the Service Program“ p. 3 - 3).
Calibration
1. Start communication via menu
File
Connect
(F1).
2. After activities on the drive:
Transfer data via menu
EEPROM
Read
(F3) from the unit
to PC.
3. After replacement of the E-Module:
Call in default data from the PC via menu
EEPROM
De-
fault
(F2). The existing values are deleted.
4. Check or input the serial number in menu
Calibration
Se-
rial Number
. Default data cannot be transferred to the device
if the serial number was not input.
5. Calibrate length via menu
Calibration
Length Calibration
(F5).
a) Insert zero point gauge in OPS slot.
b) Push drive manually to gauge and lock.
c) Start calibration.
6. Pressure calibration:
a) Open drive lock before starting pressure calibration.
b) Calibrate pressure via menu
Calibration
Pressure Ca-
libration
(F6). Insert calibration gauge (66-80N) in Ops
slot when prompted.
c) Check PWM values
Calibration point 1: Force 20 N, PWM max. 45%

Table of Contents

Related product manuals