Service Program
3
Perfusor® compact, 2.1 gb 3 - 13
C {number of the 50 ml free type} F – alarm (if the syringe is
accepted, then the syringe selection setting is not correct!)
Test for setting the free type or EEPROM:
3. Open holder, press keys 7 C {number of the 20 ml free type}
F and keys 7 C {number of the 50 ml free type} F – syringe is
accepted (if an alarm is triggered, then the syringe selection
setting is not correct!)
Press keys 7 C X X F and keys 7 C Y Y F – alarm (if the syringe
is accepted, then the syringe selection setting is not correct!)
Test for table and free type setting:
4. Open holder, press keys 7 C X X F and keys 7 C Y Y F – syringe
is accepted (if an alarm is triggered, then the syringe selec
-
tion setting is not correct!)
5. Open holder, press keys 7 C {number of the 20 ml free type}
F and keys 7 C {number of the 50 ml free type} F – syringe is
accepted (if an alarm is triggered, then the syringe selection
setting is not correct!)
Test for OPS setting:
6. Open holder, press keys 7 C 5 0 F and keys 7 C 2 0 F – syringe
is accepted (if an alarm is triggered, then the syringe selec
-
tion setting is not correct!)
7. Press keys 7 C X X F and keys 7 C Y Y F – alarm (if the syringe
is accepted, then the syringe selection setting is not correct!)
8. Reset syringe types after the test is finished!
Syringes 20 ml Free Type / Syringe Type Number
1. Open holder, press keys 7 C {number of the 20 ml free type}
F – syringe is accepted (if an alarm is triggered, then the sy
-
ringe selection setting is not correct!) Check whether the free
type was marked on the unit (please see syringe table).
2. Reset syringe types after the test is finished!