Service Program
3
Perfusor® compact, 2.1 gb 3 - 5
What to Do if (Trouble Shooting)
... the length calibration does not start?
Could communication be started successfully? Does the motor
still not start?
Then: Select Termination. Switch off pump. Repeat communica-
tion start. Switch pump on again.
... the communication to the pump is missing?
Is the service cable connection okay? Is the MFC correctly con-
nected?
Then: Select Termination. Switch off pump.
Repeat communication start. Switch pump on again.
... the communication cannot be started?
Was the setting in the File / Configuration file (COM 1 oder 2)
menu selected correctly? Is the service cable connection okay? Is
the MFC correctly connected?
... the communication starts and is then interrupted?
Then: Press the ON-key on the Perfusor compact until the sym-
bols and disappear.
... the unit does not accept any syringe after a service was car-
ried out?
Is syringe selection set to „free type“, but „free type“ was not
loaded?
Then: Set syringe selection to table / OPS or load corresponding
syringe.