Service Program
3
3 - 4 Perfusor® compact, 2.1 gb
Unit Calibration
The unit is to be calibrated (see „Unit Calibration“ ➨ p. 3 - 14) af-
ter the E-Module or the drive was replaced or the bolus rate was
changed.
Defaut Data
The Service Program contains the Default.dat file with the factory
settings of the unit. These values can be adjusted via the Syringe
or Modes menu if required.
Max. delivery rate (basal rate). . . . . . . . . . . . . . . . . . . . . 99.9 ml/h
Bolus rate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 800 ml/h
Staff call . . . . . . . . . . . . . . . . . . . . . . . . . . dynamic with Off-alarm
Alarm tone . . . . . . . . . . . . . . . . . . . . . . . . . . 0 (3 Hz interval tone)
Pressure stage. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Dianet address . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
Syringe selection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Table
Service interval. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20440 hrs.