Unit Elements
4
Perfusor® compact, 2.1 gb 4 - 11
4.10 N-Module
Designation Ord. No.
N-Module (220 -240 V) . . . . . . . . . . . . . . . . . . . . . . . . 3450 6683
N-Module (100 -120 V) . . . . . . . . . . . . . . . . . . . . . . . . 3450 6730
Exchange
1. Open housing (see „Open unit“ p. 4 - 3).
2. Remove MFC socket.
3. Pull off the N-Module connector on the A-Module (slightly
pull out the A-Module).
4. Loosen both screws (on the rear) and exchange N-Module.
5. Assembly is done in reverse order.
Note
Lay two-wire cable with mains connector behind bearing. Con-
nect mains connector correctly to the A-Module (please see fig-
ure). Do not squeeze the cable (see „Close Unit“ p. 4 - 4).
Note
The connector on the E-Module can be easily connected when the
E-Module is swivelled out (see „E-Module“ p. 4 - 10).
4.11 Housing Upper Part, Complete
Designation Ord. No.
Housing upper part incl. membrane keyboard, . . . . . . 3450 6586
carrying handle and LS-clip
Exchange
1. Open housing (see „Open unit“ p. 4 - 3).
2. Modify modules.
3. Close housing.
Note
Do not squeeze the cable (see „Close Unit“ p. 4 - 4).
Fig.: 4 - 12
N-Module