Service Program
3
Perfusor® compact, 2.1 gb 3 - 11
Procedural Instructions for Inspection after
Modifications via the Service Program
Serial Number
1. Switch on unit.
2. Start the Service Program.
3. Read out EEPROM and compare the serial number in Calibra-
tion / Serial number with the serial number indicated on the
type plate.
4. Switch device off.
Maximum Basal Rate
1. Switch on unit.
2. Insert syringe and confirm (or select), e.g. OPS 50 ml.
3. Set delivery rate 99.9 and start.
4. When the basal rate is reduced the unit triggers an alarm and
displays the maximum basal rate.
5. Acknowledge by starting again. Now the unit delivers the
maximum basal rate.
Bolus Rate (if not blocked)
1. Press the F1 key (bolus rate). The value displayed must corre-
spond to the bolus rate set.
2. Start bolus (press F and 1 simultaneously). Pump must deliver
in bolus mode and the volume infused in bolus mode is dis
-
played.
Staff Call
1. Plug MFC service connector on the MFC connector of the
unit.
2. Open syringe holder. An alarm is triggered and the LED on the
service connector flashes.
a) If „dynamic“ was set the red LED lights up for one second.
b) If „static“ was set the red LED lights up until the alarm is
acknowledged.
3. Switch device off. If “dynamic with alarm off” was set the red
LED on the service connector lights up for one second.
4. Pull off MFC service connector.