Checks after Repair
5
5 - 4 Perfusor® compact, 2.1 gb
Pressure Cut-Off
CAUTION
The limit values of the Checklist for Checks after Repair do not
correspond to the TSC values.
1. Set pressure stage 1 (key sequence F 3 C 1 F START).
2. Pump continues to deliver at 96.0 ml/h and switches off when
the specified pressure is reached.
3. Enter value in check list.
4. Set pressure stage 2 (key sequence F 3 C 2 F START).
5. Pump continues to deliver at 96.0 ml/h and switches off when
the specified pressure is reached.
6. Enter value in check list.
7. Set pressure stage 3 (key sequence F 3 C 3 F START).
8. Pump continues to deliver at 96.0 ml/h and switches off when
the specified pressure is reached.
9. Enter value in check list.
WARNING
DANGER OF INJURY! DO NOT OPEN TOGGLE AS CALIBRATION
GAUGE CAN RELEASE SUDDENLY.
10. Release calibration gauge (key sequence F 3 0) with service
connector.
Syringe Recognition
1. Insert 20 ml potentiometer calibration gauges.
The 20 ml syringe symbol is displayed.
2. Remove potentiometer calibration gauge.
3. Insert max. 50 ml potentiometer calibration gauges.
The 50 ml syringe symbol is displayed.
4. Remove potentiometer calibration gauge.