Index 01
(Master - to be added to the documentation)
Perfusor® compact, 2.1 gb 7 - 1
Sheet 1 of 2
M659 01 00 06F04 01 / 3891 1128
7 - Technical Safety Check TSC
Checklist for Technical Safety Checks – Every 24 Months
Unit: Perfusor compact infusion syringe pump
Manufacturer: B. Braun Melsungen AG
User
Observe the service manual and the instructions for use. All measured values are to be documented.
Accessories used should be included in testing. Make exclusive use of calibrated measuring equip
-
ment.
Article No. Unit No. Year of Procurement
1. Visual Inspection
Unit clean, complete, un-
damaged
Syringe fastening:
Syringe holder, axial posi-
tioner, drive head, clamp,
push-button sensor
Membrane keyboard
Battery compartment cover
and battery contacts
Unit feet
Mains lead and connector
MFC lead and connector
Holder for pole fixation, side
snap-in mechanism
2. Functional Inspection
Switch on unit.
Compare with instructions
for use: LCD self-test and
audible alarm
Compare: set delivery date
and value displayed
Check switching capability
of staff call (accessories)
Switch on unit in battery
mode and check self-test
Note
If „Battery discharged“ is dis-
played:
Charge battery or replace bat-
teries and repeat test.
Compare status display
000A or 000„b with bat-
tery cells used.
Check push-button sensor
alarm
Check positive locking sen-
sor alarm
3. Pressure Cut-Off
(alternatively with manometer or check
gauge)
With manometer and 50 ml sy-
ringe:
Pressure stage 1 <0.6 bar
Pressure stage 2 <0.9 bar
Pressure stage 3 <1.2 bar
With check gauge current step
Order - No. 0770 1616:
Pressure stage 1 <40 N
Pressure stage 2 <59 N
Pressure stage 3 <75 N
CAUTION
Danger of injury: Remove check
gauge only when released.
4. Syringes
Syringe selection
OPS
internal
EEprom
Syringe table readable
Yes
No
Syringe recognition
Manufacturer (code) used
20 ml
50 ml
_____
_____
(Part 1 of 2)