The procedure continues automatically with the priming phase:
4. Once the saline reaches the drain bag, increase the level of
the priming fluid in the venous drip chamber, while the pump
is turning, using the adjustment knob.
The level should be well above the air detector head.
5. When the blood pump has stopped, turn the dialyzer (red
end up).
CAMP/MG208/20-0006 AK 98 Independent Home Dialysis System 14
Priming the dialysis circuit – cont’d