CAMP/MG208/20-0006 AK 98 Independent Home Dialysis System 29
Emptying the blood lines
1. Close clamps on the saline bag.
2. Disconnect arterial line from the Rinse back line.
3. Connect venous line to drain bag.
4. Remove both arterial and venous blood lines from
the priming detector.
5. Turn the dialyzer so the blue end is up.
6. Open arterial line clamp.
7. Start Blood pump.
8. When the circuit is empty, stop Blood pump.