EasyManua.ls Logo

BD Alaris GP - Alarms: Specific Alarm Conditions; Specific Alarm Conditions

BD Alaris GP
58 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Alaris™ GP (Guardrails™) Volumetric Pump
Alarms
BDDF00535 Issue 1
33/56
Display Infusion Status Cause(s) Action
NO FLOW Infusion stopped Flow sensor detects no flow.
Check flow sensor.
Check fluid level in container.
Ensure all clamps above Pump are
open.
Ensure drip chamber is half filled.
Ensure that the bag spike is inserted
correctly.
Check flow sensor is clean.
FLOW ERROR Infusion stopped
Gross difference between
detected drops and expected
amount of drops.
Clamp infusion set using roller clamp.
Check flow sensor.
Check fluid level in drip chamber.
FLOW ERROR (In secondary
infusion mode only)
Infusion stopped Unexpected drops detected.
Hang secondary container above
primary.
Check drops are from secondary
container when infusing.
Flow sensor disconnection is
recommended.
FREE FLOW Infusion stopped Uncontrolled flow possible.
Clamp infusion set using roller clamp.
Remove Pump from use.
BATTERY EMPTY Infusion stopped
The internal battery is
exhausted. The Pump will
automatically switch off in the
immediate future.
Connect to power supply immediately
or switch Pump off.
SAFETY CLAMP Pump on hold
Safety clamp broken or
missing.
Clamp infusion set using roller clamp.
Replace infusion set.
Investigate and correct set loading.
SET MISLOAD Pump on hold Set loaded incorrectly.
Clamp infusion set using roller clamp.
Investigate and correct set loading.
FLOW SENSOR DISCONNECT Infusion stopped
Flow sensor unplugged
during infusion.
Check / replace flow sensor or set VTBI.
WRONG SET Pump on hold Safety clamp not detected.
Clamp infusion set using roller clamp.
Check set and close door.
Replace infusion set (If necessary).
DOOR CLOSE INCOMPLETE Pump on hold
Safety clamp in non-occluded
position with door open or
obstructed.
Clamp infusion set using roller clamp.
Investigate and correct set loading.
Close door.
DO NOT USE
Pump on hold /
infusion stopped
Internal error has occurred. Remove Pump from use.
LEVER OPEN Infusion stopped Door lever is open
Check door lever.
Check lever hooks.
Check lever is not obstructed, if so, free
obstruction.
VTBI DONE Infusion stopped Intended VTBI completed. Set new VTBI or clear VTBI.
SET CLOCK Pump on hold Date / time not set.
Qualified Service Personnel must set
date / time.
Press CANCEL softkey to continue.
SET SERIAL NUMBER Pump on hold Serial number not set.
Contact Qualified Service Personnel to
set the serial number.

Table of Contents

Related product manuals