6.2.5 Channel descriptions for the process image
Value
Measured value
description alarm
message operating
message
Note
0
1 (0x01) Insulation fault
2 (0x02) Overload
3 (0x03) Overtemperature
4 (0x04) Failure line 1
5 (0x05) Failure line 2
6 (0x06) Insul. OP light Insulation fault operating theatre light
7 (0x07)
8 (0x08) Distribution board failure
9 (0x09) Failure oxygen
10 (0x0A) Failure vacuum
11 (0x0B) Anaesthetic gas
12 (0x0C) Compressed air 5 bar
13 (0x0D) Compressed air 10 bar
14 (0x0E) Failure nitrogen
15 (0x0F) Failure CO2
16 (0x10) Insulation UPS Insulation fault UPS
17 (0x11) Overload UPS
18 (0x12) Converter UPS
19 (0x13) UPS fault
20 (0x14) UPS emergency operation
21 (0x15) UPS test run
22 (0x16) Failure air conditioning
23 (0x17) Batt.op. OP-L Battery-operated operating theatre light
24 (0x18) Batt.op. OP-S Battery-operated Sat OP light
25 (0x19) Fail.norm.supply Line normal power supply
26 (0x1A) Fail.safet.supply Line safety power supply
27 (0x1B) Failure UPS Line additional safety power supply
28 (0x1C) Ins.safety supply
29 (0x1D) Fail.N conductor
30 (0x1E) Short dist. panel Distribution board short-circuit
31 (0x1F)
32 (0x20)
33 (0x21)
34 (0x22)
35 (0x23) Standby function (Measuring function switched off (standby))
36 (0x24)
37 (0x25)
38 (0x26) Batt.op. UPS Battery operation, special safety power supply
39 (0x27) Phase sequ. left