BD/ED/FD (E2) Service Manual
State: 01/2002 Created: 06/2002 / Jochen Tussinger
13
4.3 Calibration Procedure “One Point Calibration”
Press the Buttons and for 5 sec.
simultaneously.
After the 5 sec. you’ll see C211 in the display.
Don’t use this configuration for the One Point
Calibration. You have to use the configuration
C212.
Press the Button to go to configuration C212.
Set the measured value with the Buttons and
like displayed on your reference measurement
system.