PATIENT SAFETY
PATIENT SAFETYPATIENT SAFETY
PATIENT SAFETY-
--
- C
C C
CONT.
ONT.ONT.
ONT.
6
66
6
It is not recommended that entrainment be turned on or off while ventilating. However, if this
should become necessary, then it should be done during the expiratory phase of the breathing
cycle. Be aware that this will change the flow.
Approximately twenty minutes of operation will remain after a low battery alarm assuming a
properly maintained battery in good condition.
The battery should be replaced at least every two years. Only use batteries supplied by Bio-Med
Devices, part #PRT4467.
Prior to disposal of any component, with particular attention to the battery and PCB, check with
your local controlling authority for disposal regulations.
When first turned on, the unit displays “Bio-Med Devices, Inc.” until the initialization process is
complete. If this does not clear, submit unit for service.
Be aware that the SIMV rate is also the backup rate which is the rate the patient will get in the event
of apnea.
While scrolling a parameter, some hesitation may be observed.