APPENDIX C
SUMMARY OF PARAMETER RANGES / ACCURACIES
SUMMARY OF PARAMETER RANGES / ACCURACIESSUMMARY OF PARAMETER RANGES / ACCURACIES
SUMMARY OF PARAMETER RANGES / ACCURACIES
PARAMETER RANGE RESOLUTION OF DISPLAY ACCURACY
___________ __________ _________________________ ________________
Rate 5- 150bpm .1 bpm below 10; 1 bpm above 10 ±10%
Tidal Volume 5- 2500ml 1ml below 200; 10 ml above 200 ±10%
1
Flow Rate 1- 120 lpm .1 lpm below 10; 1 lpm above 10 ±10%
2
Peak Pres 0- 120 cmH
2
O 1 cmH
2
O ±3% FS
PEEP Pres 0- 35 cmH
2
O 1 cmH
2
O (above baseline) ±3cm H
2
O
Pressure Trigger -10 to -0.2cm .1 cmH
2
O below 3; 1cm above 3 ±1cm H
2
O
Press Supp. 0- 50 cmH
2
O 1 cmH
2
O (above baseline) ±3cm H
2
O
SIMV Rate .6- 50bpm Same as Rate ±10%
O
2
sensor 18- 100% 1% ±3% FS
EXHTV 50– 3200 ml 1 ml below 200; 10 ml above 200 ±15%
3
EXHMV 0-45 L 0.1 L below 10 L, 1 L above 10 L ±15%
1
Accurate from 100-2500 ml
2
Accurate from 1-120 lpm ±10% of setting or 1 lpm, whichever is greater. With entrainment on, flow
accuracy from 5 - 100 LPM is ±10% or 1 LPM, whichever is greater, and with a flow of 100 LPM and over,
flow accuracy is within 15% of displayed flow. No accuracy is claimed below 5 LPM.
3
Readings below 100 ml may be considered relative rather than absolute values.