IX. MAINTENANCE AND SERVICE- cont.
47
4747
47
PRESSURE SUPPORT
PRESSURE SUPPORTPRESSURE SUPPORT
PRESSURE SUPPORT
(MAIN MENU)
Set PRESSURE SUPPORT to 10 cmH
2
O and initiate breaths using inspiratory effort. Verify that the high-
pressure signal to the exhalation valve is terminated when the circuit pressure equals 10 cmH
2
O above
PEEP during spontaneous breaths only, and that the gas flow remains on for the duration of the set
inspiratory time.
PRESSURE TRIGGER
PRESSURE TRIGGERPRESSURE TRIGGER
PRESSURE TRIGGER
(MAIN MENU)
Tee a calibrated low-pressure analog gauge into the airway pressure line of a patient circuit.
Set Pressure Trigger to -1.0 cmH
2
O and verify the CROSSVENT triggers a breath when a pressure of -1
cmH
2
O is applied to the circuit.
Repeat for -5 cmH
2
O.
CPAP FUNCTIONS (SET TO CPAP MODE)
MANUAL
MANUALMANUAL
MANUAL
Press the MANUAL key and verify a pressurized breath is delivered and MAN is displayed in the lower left
key during this breath.